ECS05 Gradient Analytical System

Highly effective HPLC system containing a UV-VIS detector, a column oven, an analytical pump, a gradient box, and an autosampler.

Detector with eight channels in range of 200-800 nm and very low noise level offers also high-speed scanning with speed up to 100 Hz. Communication of units is through RS232 or Ethernet(LAN). An additional PC A/D converter card is not required for this system. The system contains autosampler and Clarity software. All units are supported in Clarity software from installation. 

A wireless mouse and keyboard set (P/N: EPC-0028), PC (P/N: EPC-0026) and a computer monitor (P/N: EPC-0027) can be purchased with the system.

Main parts

ECDA2800 UV-VIS PDA DETECTOR
Analytical Diode Array Detector measures at eight wavelengths simultaneously or performs a full scan with a speed up to 100 Hz. 3D support in Clarity SW. Noise level ±5x10-6 AU. Four analog outputs. Two special mathematical channels are available. AD 05 analytical cell is included.
 
ECO2080 COLUMN OVEN WITH HEATING AND COOLING
This column oven allows precise temperature stabilization in the range of 0-80°C. A big compartment room can accommodate up to 3 columns with lengths up to 25 cm. Peltier technology offers the possibility of heating and cooling independently from room temperature.
 
ECP2010 ANALYTICAL HPLC PUMP
This isocratic pump works with Gradient Box ECB2004B as a quaternary gradient pump. The flow rate range is 0.02-10.00 ml/min and the pressure limit at 40 MPa. Unit software includes a new learning algorithm for pulsation suppression and many testing and diagnostic functions.
 
ECB2004B GRADIENT BOX WITH DEGASSER
The box is suited for handling liquids in ECOM gradient systems. It accommodates a container for solvent bottles, a built-in high-efficiency vacuum degasser, and four-way gradient valve. Up to four solvent lines may be degassed simultaneously by one unit. A four-way gradient valve is controlled from the pump ECP2010.
 
AUTOSAMPLER L-3320
Automatic autosampler ensures excellent accuracy and linearity of sample injection. Very low sample residue contamination, μl pick-up mode achieves zero sample loss. Automated self-checking. Rheodyne valve ensures reliable operation.
 
CLARITY CHROMATOGRAPHY SOFTWARE
This system is fully controlled by Clarity Chromatography software, which is also a part of delivery. Clarity software complies with CFR21GLP requirements.
Detection● UV-VIS PDA up to 800 nm ● 8 channels simultaneously ● Scan: full spectrum, high speed up to 100 Hz
Flow Rate (ml/min)10 ml/min
Main System Parts● Detector: (ECS04) UV-VIS, (ECS05) UV-VIS PDA ● HPLC pump ● Gradient box with degasser ● Column oven with cooling and heating ● Autosampler ● Analytical injection
Max. Pressure40 Mpa/ 5802 psi
No. of Solvents4
TypeHPLC Analytical Gradient System

Generate catalog for :

SYSTEM CONTAINS 
 
PcsDescriptionP/N
1ECDA2800 Analytical UV-VIS PDA DetectorANW0000X
1ECB2004B Gradient box with degasserABA0000X
1ECP2010 Analytical HPLC pumpACA0000X
1ECO2080 Column oven with heating and coolingAOA0000X
1Autosampler L3320 for 2x48 vialsARA03320
1Clarity – Station for 1 instrumentAP000C50
1Clarity module for HPLC controlAP000A24
1Clarity SW module for AS controlAP000A26
1Module for PDA data Clarity (SW)AP000A29
1Accessories for analyt. system series 2000ASA00180
1Accessories ECOM for AS Rigol L3320/L3320HARA00180
Note: Column and computer are not included. PC system has to be equipped for handling 3 serial RS232 or 1 Ethernet port at least. Check also Clarity software requirements.